Medical
Device
& QA Consultancy
CUSTOMER SUPPORT SERVICES
Changing to an Alternative Authorized Representative
Definition of Authorized Representative
Definition of Authorized Representative, as amended by the IVD
Distributor acting as Authorized Representative
Need for an Authorized Representative
Reporting path between the Manufacturer and the Competent Authority
Role of the Authorized Representative
Selecting an Authorized Representative
Return to our HOME PAGE E-mail: bernardt@mdqaconsultancy.co.uk
Introduction
There
are many misunderstandings and myths surrounding exactly who represents the
manufacturer in Europe, when the manufacturer himself is based outside the
European Community. The exact role and the responsibilities of the
‘manufacturers representative’ are not clear either, therefore one of the
objectives of this chapter is to clear up this confusion so that manufacturers
based outside the European Community can make informed decisions and ensure
that their best interests are served at all times.
The
term ‘authorized representative’ has different meanings and therefore different
responsibilities under other directives. For example the Machinery, Low Voltage
and certain other directives mention the role of the authorized representative,
but these are outside the scope of this publication. But in the context of the
Medical Device directive 93/42/EEC, the authorized representative has a special
and particular role to play.
The
term ‘manufacturer or his authorized representative’ is referenced 18 or so
times in the medical device directive itself, but astonishingly, it is not
defined in Article 1 of the directive. Article 1 of the directive is the
section which deals with definitions. Various terms referred to in the
directive are defined, such as ‘the manufacturer’, ‘placing on the market’ and
most importantly a ‘medical device’, but a definition of the ‘authorized
representative’ has been omitted.
For
completeness, the following table lists all the annexes and articles of the
directive, where a reference is made to the ‘authorized representative’.
Articles and
Annexes of the medical device directive 93/42/EEC that are relevant to the
Authorised Representative
· Articles 8 - Safeguard Clause issues
· Article 10 - Adverse Incident Reporting responsibilities
· Article 11 - Co-operation with Notified Body in Conformity
Assessment issues.
· Article
14 - Registration of Person Responsible for "Placing on the Market".
· Article 15 - Clinical Investigation issues
· Article 16 - Co-operation with Notified body on time
limits etc
· Article
18 - Co-operation with Notified Body and Competent Authority in "Wrongly
Affixed CE Marking" issues.
· Article
19 - To represent the Company 's interests in decisions of Refusal or
Restriction
· Article
21 - Represent the Company in matters concerning the Repeal & Amendment of
related Directives.
· Annex
I - To ensure that the Company meets all it's responsibilities under ER #'s
13.3 and 13. 6., Labelling and Instructions for Use.
· Annex
III & IV - To represent the Company's interests, where appropriate.
· Annexes
II, IV, V, VI, VII and VIII - To represent the Company's interests in the
appropriate Conformity Assessment routes
· Annex X - To represent the Company's interests.
The
basic need to have an ‘authorized representative’ arises because, where a
manufacturer is based outside the European Community, the Competent Authorities
need to be able to contact an entity or person who is based within the European
Community, and who acts on behalf of the manufacturer.
There
are three terms within the directive, which have an impact when the
manufacturer is based outside the European Community. These are the ‘authorized
representative’, ‘the person responsible for placing products on the market’
and ‘the importer’.
Where
manufacturers are based outside the European Community, there is a sound
argument for differentiating between the regulatory (including product safety)
issues and the marketing issues involved in selling product in Europe.
Marketing people generally have sales targets and market share issues to keep
them occupied, whilst compliance with the regulatory issues requires a separate
and special focus, to keep the Company on the right side of the regulation.
Most, but not all, companies do differentiate between the two issues and
appoint separate entities to execute the two functions. However, it must be
said that the marketing and the regulatory issues can be executed by one
entity, if the manufacturer so wishes.
Article
14 is entitled 'registration of persons responsible for placing devices on the
market' and is the directive's requirement for the manufacturer of class I
devices and procedure packs to inform relevant Competent Authorities of certain
information. According to article 14, the details required from the
manufacturer are;
· the address of the
registered place of business and
· a description of the
devices
This
information has to given to the "competent authorities in the member state
in which the person responsible has their registered place of business".
Therefore, this implies that registration only needs to be carried out once. In
reality, some competent authorities are requiring registration of all devices
of all Classes, not just Class I.
The
article goes on to state in clause 2 that where the manufacturer is based
outside Europe, that the exact same requirements apply to the designated [by the
manufacturer] person responsible for marketing.
Clearly,
a hierarchical structure is emerging, which can be summarised as follows;
MANUFACTURER Designates an
entity to act on their behalf when addressing authorities and bodies
The 'entity' can be either
the person responsible
for marketing the devices the authorized
representative of the manufacturer established within the community the importer established
within the community
or or
The manufacturer and the
manufacturers representative's name on the Label
Accurate
labelling of all medical devices is of the utmost importance, not least because
the directive states that a product recall could become necessary because of
"...any inadequacy in the labelling or the instructions for use..."
(Annex VII Clause 4 ( i ). The labelling criteria for all medical devices is
set out in Annex I (Essential Requirements) of the directive, under Essential
Requirement number # 13.3. The particular requirements for the manufacturer and
the 'manufacturers representative' are contained in sub requirement (a) of 'essential requirement' number 13.3. This
'essential requirement' reads as follows;
"...the
label must bear the name or trade name and address of the manufacturer. For
devices imported into the community, in view of their distribution within the
community, the label, or the outer packaging, or the 'instructions for use',
shall contain in addition the name and address of either;
· the person responsible [for
marketing the devices], or of
· the authorized
representative of the manufacturer established within the community, or of
· the importer established
within the community,
as
appropriate...".
Summary of the requirements
of the 'Label'.
Labelling
is the collective term which refers to;
· the actual Label itself,
which would normally be affixed to the medical device
· the 'outer packaging'
· the 'Instructions for Use'
· any 'product inserts'
The
medical device directives requirements for the Label are set out in Essential
Requirement # 13.3. The particular requirement for information regarding the
manufacturer is contained in clause (a) of Essential Requirement # 13.3.
Where
the manufacturer is located outside the European Community, the clause requires
that;
"...the
label, the outer packaging or the instructions for use, shall contain the name and
address of either the person responsible for marketing the devices or the
authorized representative of the manufacturer established within the Community
or of the importer..."
The
above requirement can be simplified as follows;
PRODUCT LABEL INSTRUCTIONS FOR
USE PRODUCT INSERTS OUTER PACKAGING
The or the or
any or the
must contain either;
(1) - The NAME OF THE PERSON RESPONSIBLE FOR MARKETING
or;
(2) - The AUTHORIZED REPRESENTATIVE OF THE MANUFACTURER
ESTABLISHED WITHIN THE COMMUNITY
or;
(3) - The IMPORTER ESTABLISHED WITHIN THE COMMUNITY
Amended definition of
‘authorized representative’.
As previously
stated, the medical device directive 93/42/EEC does not define the role or
responsibilities of the authorized representative. The medical device directive
entered into force in June 1995, but it was not until 3 years later that the
role of the authorized representative was actually defined.
The
entry into force of the In vitro diagnostics directive (IVD) (98/79/EC) in
December of 1998, did at last provide a definition of the ‘authorized
representative’. In Article 21 ( j ) of directive 98/79/EC, which
retrospectively amends the medical device directive, the following definition
is given;
‘authorized
representative’ – means any natural or legal person established in the Community
who, explicitly designated by the manufacturer, acts and may be addressed by
authorities and bodies in the Community instead of the manufacturer, with
regard to the latter’s obligations under this directive.
Role of an authorized
representative
It
is very important for any manufacturer to clearly define in their Standard
Operating Procedures (SOP's) covering Post Sales Issues, the exact
responsibilities and reporting paths for all activities. FIG 1 illustrates
typical reporting paths. The following Post Sales Issues could all involve the
authorized representative, dependent on the manufacturers preference and just
how the system had been set up;
•
Customer
Complaint Handling
•
Adverse
Incident Reporting
•
Product
Recall
•
Post
Market Feedback
Of the above issues, the
authorized representative's involvement in Post Market Feedback and Product
Recall would be optional. However, the authorized representative would
certainly be involved in Customer Complaint Handling and Adverse Incident
Reporting. Not least because the authorized representative's name and address
would be on either the Label itself or the IFU, or possibly both. Any user then
requiring to make a complaint, would naturally direct the complaint to the
entity displayed on the Label or in the Instructions for Use (IFU).
So, the manufacturer would
need to define and document in their procedures, the reporting path for
Customer Complaints, should a complaint be received by the authorized
representative.
The contract between the authorized
representative and the manufacturer should clearly and unambiguously assign
responsibilities for all activities.
Complaints could also be fed
back to the manufacturer or the authorized representative by way of
distributors. Therefore, a contract between a distributor and the manufacturer,
besides dealing with marketing issues, should make clear the reporting paths
and responsibilities for these issues. FIG 1 illustrates the main issues to be
considered, along with typical reporting paths.
There are two possibilities
regarding the reporting of reportable incidents. Firstly, that when the
authorized representative receives a Customer Complaint that meets the
Reporting Criteria, the authorized representative then reports direct to relevant Competent
Authorities, without prior consultation with the manufacturer. Parallel
reporting back to the manufacturer would be essential in this scenario.
The second possibility
is that the authorized representative
feeds back all Customer Complaints to the manufacturer so that the
manufacturer, through their Corrective Action system, can make a judgement on
whether or not the reporting criteria has been met. The reporting path to the
competent authority could then be either direct or through the authorized
representative. If the former is the case, then parallel reporting would be
necessary to the authorized representative, to keep him aware of the
situation.
In most medium to large
sized companies, where a mature Quality Management System would be in place,
the direct reporting path to the competent authority, with parallel reporting
to the authorized representative, would be the preferred option. This would
enable the manufacturer to exercise complete control over the judgement as to
whether or not the reporting criteria had been met, and whether or not the
consequent reporting to relevant competent authorities was necessary.
The former option, where the
authorized representative exercises the control and judgement on which incidents
to report, might be appropriate for a very small company, which needed to rely
on the expertise of a third party, where regulatory issues are concerned.
If the former option is
chosen, it is imperative that the manufacturer is fully involved every step of
the way, as the manufacturer has the final responsibility for all issues.
Therefore, it must be
restated that the responsibilities and reporting paths of FIG 1 need to be
established, documented and if necessary, included within contracts or
agreements.
Reporting criteria
There are two kinds of
reportable incident;
Adverse Incident - an incident that caused the death or serious deterioration in the
health of a patient, user or other person,
Near Incident - an incident that might have caused the death or
serious deterioration in the health of a patient, user or other person,
It is important to
differentiate between the above two types of Reportable Incident, because there
are different reporting times. Ten (10) days for an Incident and thirty (30)
days for a Near Incident.
These definitions should be included within the
manufacturers Corrective Action procedures and the persons responsible should
be trained accordingly. Records should be kept of this training.
In accordance with FDA 21 CFR 820.25 requirements,
these persons should additionally be trained to understand the consequences of
a failure to implement the procedures correctly.
Responsibilities of the
authorized representative
The authorized representative's responsibilities
could be separated into two categories; namely 'minimum regulatory
responsibilities' and 'additional responsibilities'.
Minimum regulatory
responsibilities
The minimum service that any authorized
representative offers should include the following issues. Referring to FIG 1,
the authorized representative would be involved in receiving Customer
Complaints from Customers themselves or from distributors, and feeding back
that information into the manufacturer's Corrective Action system.
Additionally, registering the manufacturers products
with appropriate Competent Authorities and holding a copy of the Technical
File, to be provided to a Competent Authority on request, are important issues
to be addressed. If the authorized representative holds a copy of the Technical
File, then the manufacturers quality management system document control needs
to extend to the Technical File copy held by the authorized representative.
This change notification requirement ought to be part of the agreement between
the manufacturer and the authorized representative.
In cases where the authorized representative does
not hold a copy of the Technical File, then the agreement needs to ensure that
the manufacturer will supply a copy of the Technical File to either the
authorized representative or/and the Competent Authority, within 7 days of a
request.
Significant change
notification
Where significant changes occur to either the
manufacturers quality management system or product range, (FIG 1 block C) then provision
needs to be made to inform the authorized representative when such changes
occur. Again, this requirement should be part of the agreement between the
authorized representative and the manufacturer.
One of the most important responsibilities of the
authorized representative is to 'act and to [may] be addressed by
authorities and bodies in the Community instead of the manufacturer, with
regard to the latter’s obligations under this directive', (see 'amended definition of authorized
representative' earlier in this chapter).
This is an important function and can involve simply communicating
verbally or in writing with Competent Authorities or in extreme cases, making
an actual visit. Issues that may fall within this scope could be for example, device
classification issues (where an error in classifying a device had been
discovered), issues involving product recall or possibly where a product's
performance had been called into question.
Another very important issue is that of product
recall. Such an issue would generally be controlled from the manufacturers
facility outside Europe, but it would be sensible to involve the authorized
representative, not least because of his closer proximity to the installed base
of product.
One final consideration is where the manufacturers
products are of an 'Implantable' nature. An Implantable medical device,
according to Section I of Annex IX (Classification criteria) of the medical device directive 93/42/EEC is
as follows;
Any device which is intended;
-
to be totally introduced into the human body or,
-
to replace an epithelial surface or the surface of the eye,
by surgical intervention which is intended to remain
in place after the procedure.
Any device intended to be partially introduced into
the human body through surgical intervention and intended to remain in place
after the procedure for at least 30 days, is also considered an Implantable
device.
In
addition to the requirements that EN 46000 imposes on all medical devices,
there are particular additional requirements for 'Implantable Medical Devices'.
One requirement in particular is Clause 4.15.6 which deals with 'Delivery' of
product. The requirement is as follows;
"The [manufacturer] supplier shall require that
any authorized representative maintains records of distribution of medical
devices and that such records are available for inspection"
Optional responsibilities
An authorized representative service may also
include the following issues. Referring to FIG 1, the authorized representative
may also be involved in researching Clinical, Medical, Regulatory or Standards
requirements and feeding these back to the manufacturer, as part of the
agreement between the manufacturer and the authorized representative.
Additionally, the authorized representative may be
involved in implementing the manufacturers Post Market Feedback system. This is
the means by which the manufacturer evaluates the performance of their product,
in the post sales phase, in accordance with a generic documented procedure. This
procedure would incorporate the following elements;
· would be product based
· would be invoked for a
particular purpose, for example:
-
to
verify the design of the product
-
to
validate a new intended purpose for an existing device
-
to
validate customer satisfaction,
-
to
meet regulatory and vigilance requirements
· would have the data sources
defined, to include proactive and reactive sources,
Examples of reactive sources:
- Customer Complaints,
Service Reports, Returned Product etc
Examples of proactive
sources:
- Distributor and/or
Customer questionnaires, Marketing Department data, information on competitor
or similar products, etc
· would provide for review of
the gathered data,
The review should be by a designated person and
documentary evidence of that review should be recorded.
· determination of necessary
resulting actions.
Once the data had been gathered, reviewed and
analysed, the final step in the process would be to determine and implement any
necessary resulting actions. These might include;
- Changes to the design of
the product,
- Significant Change
notification, if necessary,
- Reporting of an incident which meets the reporting
criteria.
- Retraining of personnel
- Changes to quality management system procedures or
work instructions.
The procedure should also make provision to review
the process at defined intervals. In addition, the procedure should provide for
a review of the Post Market Feedback System itself, to ensure it's continuing suitability
in meeting the 'purpose ' for which the procedure was invoked in the first
place.
Summary of Post Sales Issues involving the
Authorized Representative
CHANGE CONTROL SYSTEM
![]()
![]()
![]()
![]()
PRE
SALES ISSUES POST SALES
ISSUES

![]()



![]()
Copyright: Medical Device & QA Consultancy - http://www.mdqaconsultancy.co.uk FIG 1
Selecting an authorized representative
The manufacturer has to
determine his basic needs before committing to a particular authorized
representative. Some authorized representative's offer a very wide range of
authorized representative Service. Others offer a service far in excess of that
required by the directive. Generally, the more wide ranging the service, the
higher the fee.
The considerations are
whether the manufacturers needs require an authorized representative to provide
just the minimum regulatory requirement covering Post Sales Issues or whether
additionally Pre Sales Issues such as design of new product, advice on changing
regulatory and standards requirements etc need to be included.
Typical authorized representative service
As a minimum, any authorized
representative service should include involvement in all the Post Sales Issues
of ;
· Incident Reporting,
· Product Recall,
· Complaint Handling, and
· Post Market Feedback.
Additionally, the chosen
authorized representative should have the competency to be able to
'act and to [may] be
addressed by authorities and bodies in the Community instead of the
manufacturer, with regard to the latter’s obligations under this directive', (see
'amended definition of authorized representative' earlier in this chapter), as
and when required, for example;
· Registering Class I devices
with relevant competent authorities
· Notifying Significant
Changes to appropriate bodies
· Classification or product
performance issues
· Hold the technical file for inspection by
the authorities or be the conduit, by which the
Technical File is delivered to the authorities, as required
It is not absolutely
necessary for the authorized representative to be totally familiar with the
manufacturers particular products, but an authorized representative with
experience of a wide range of products, is more likely to be able to serve the
manufacturers best interests.
An important area where an
experienced authorized representative would be of benefit to a manufacturer, is
where Class I devices were involved. Because of the Self Certification nature
of Class I devices, where the product is not a measuring device and is not
supplied sterile, it is important to apply the Classification procedure effectively,
because there is no third party verification of either the Technical File
content or the appropriateness of the chosen conformance route.
In order to provide an
effective service, an authorized Representative should also be able to advise a
manufacturer on all of the above issues. To be able to be fully effective, the
authorized representative should have some means of keeping abreast of current
developments and to be aware of new interpretations of existing requirements,
as they occur.
For example, the author of
this chapter offers two kinds of authorized representative service. A Standard
Service covering the basic regulatory Post Sales Issues mentioned above and a
tailor made Premium Service covering additionally any Pre Sales Issues such as
Design Control, Product Safety and New Product Development, that are of
interest to the manufacturer.
Changing to an alternative authorized
representative
A manufacturer may decide to
change their authorized representative for a variety of reasons. Their
authorized representative may cease to trade or the manufacturer may not feel
that they are receiving value for money or the manufacturer may simply feel
that their best regulatory interests are not being professionally looked after.
Either way, the task of choosing an alternative authorized
representative is somewhat similar to changing Notified Bodies; it can be
achieved quite simply and effectively. The main consideration would be the
termination clause in the Contract or Agreement between the manufacturer and
their existing authorized representative.
Implications of changing authorized representative
The following areas need to
be considered;
· Authorized representative's
name on Label and or Instructions for Use and/or sales Literature
· Significant Change
notification
· Post Sales procedures
· Manufacturers indemnity
insurance
· Might be part of
distributor network
The authorized
representative's name and address could be on any or all of the Label itself,
Instructions for Use, Product Inserts, as well as Sales and Marketing
Literature. These would all have to be changed to reflect the new authorized
representative service being used. Product already on the market, with the
previous authorized representative's details would not need to be
retrospectively changed to reflect the new authorized representative. However,
consideration would need to be given to the fact that product was on the
market, with the name and address details of an entity who was no longer the
authorized representative of the manufacturer.
The change of authorized
representative would be a 'Significant Change' (block C of FIG 1) and would
therefore require notification to the manufacturers notified body and also to
relevant competent authorities, where Class I devices had been registered.
Some authorized
representative's offer two kinds of Service.
An authorized representative
who had a wide experience of Medical
Device Compliance strategies would be able to offer both Post Sales Regulatory coverage and Pre Sales
help and advice.
Medical
Device & QA Services offers a painless and seamless transition for those
Device Manufacturers who have identified the need to move from their existing Authorized
Representative to a New One. Procedures to enable the transition to be both
seamless and simple are available as part of our Authorized Representative
Package and Unlimited assistance is given on all related matters. Please e-mail
bernardt@mdqaconsultancy.co.uk for a quotation or telephone
+44 161 980 4310 to discuss your requirements.
Summary of considerations
when selecting an authorized representative
As discussed previously, the Importer or Distributor may act as the Authorized Representative of any Company that does not have a permanent base within Europe. Some Companies do appoint their Distributor to act as their Authorized Representative, but others realize that the Sales and Marketing responsibilities, for example;
·
Identifying new Customers
·
Increasing Market Share
·
Identifying areas into which the Company can diversify
·
Ensuring that existing Customers remain satisfied and loyal
· plus
a multitude of other Sales and Marketing issues….
Distributor acting as the authorized representative -
The main consideration is whether or not an existing distributor of a
manufacturers products can effectively fulfil the regulatory responsibilities
alongside their marketing activities. Only the manufacturer can decide, taking
full account of the issues already discussed and the following;
· how long has the prospective authorized
representative been in the health care industry ?
· do they have a demonstrable track record ?
· can they supply references from existing
manufacturers for whom they are supplying the same service ?
· can they demonstrate sufficient knowledge of
the medical device directive (93/42/EEC) itself ?
· can they advise accurately on all the Regulatory issues of Labelling,
including The Label itself, Product Inserts, Inner and Outer Packaging,
Instructions for Use and Advertising etc etc ?
· can they demonstrate sufficient knowledge of
the ISO 9000 series of standards ?
· can they demonstrate knowledge of the
additional requirements of the EN 46000 and ISO 13485 series of standards,
including the new ISO 9001:2000 ?
· are they prepared to be responsible for the
additional Implantable requirements, if appropriate ?
· are they committed to continuous improvement
of the service that they provide ?
· are they committed to keeping abreast of
developments and changes in the regulatory requirements ?
· can they provide advice and guidance on
parallel issues such as the Packaging and the Packaging Waste Directive,
Electrical and Electronic Waste Directive and the ISO 9000 year 2000 revisions
etc ?
· can they help organise or control
distributors, if required.
· can they advise on effective implementation
of all Regulatory Procedures, including Adverse and Near Incident Reporting,
Advisory Notice Issue (including Product Recall), Post Market Feedback,
Complaint Handling and Significant Change Notifications ?
· are they prepared to participate fully, if
required, in all Post Sales Issues (Post Market Feedback, Incident Reporting,
Product Recall and Complaint Handling) ?
One final consideration
Currently, there are different requirements in some
European Countries, concerning the registration procedures of Class I devices and
Systems and Procedure Packs. Therefore, the entity (distributor, importer or
authorized representative) that the manufacturer has designated to 'act
[on the manufacturers behalf] and [to] be addressed by authorities and bodies
in the Community instead of the manufacturer, with regard to the latter’s
obligations under this directive', (see 'amended definition of
authorized representative' earlier in this chapter), needs to be cognisant of
the differing Country specific requirements.
To reiterate, Article 14 of the medical device
directive 93/42/EEC requires the registration of Class 1 devices. But, in
practice some competent authorities require registration of all Classes of
device. The safest way for the time being is to register all devices with the competent
authority of each Country to which product will ship. When the IVD directive is
in force, then it will only be necessary to report to one competent authority.
Paragraph 3 of Article 14 of the medical device
directive concerns sharing of details by competent authorities, and is
reproduced here for information;
3. The Member States shall on request
inform the other Member States and the Commission of the details referred to in
paragraphs 1 and 2 [namely, the address of the registered place of business
and a description of the devices].
This was an early attempt by the European Commission
to establish a European wide database of medical devices of all classes that
had been placed on the market.
However, this extremely useful concept, which would
eventually preclude the need to report to multiple competent authorities, did
not progress beyond the concept stage. Again, it was left to the publication of
the IVD Directive to map out the way forward.
When the IVD Directive comes into force in June
2000, an appendix to Article 14 titled "European data bank" (Article
14 (a)) will apply to the medical device directive 93/42/EEC. The text is
reproduced here;
1.
Regulatory data in accordance with [ the medical device directive
93/42/EEC] shall be stored in a European database accessible to the competent
authorities to enable them to carry out their tasks relating to [ the medical device directive 93/42/EEC] on
a well informed basis.
The data bank shall contain the following;
(a)
data relating to registration of manufacturers and devices in
accordance with article 14 [of the medical device directive 93/542/EEC].
(b)
Data relating to certificates issued, modified, supplemented,
suspended, withdrawn or refused according to the [conformance] procedures laid
down in Annexes II to VII.
(c)
Data obtained in accordance with the vigilance procedure, as defined in
Article 10 [of the IVD directive 98/79/EC].
2.
Data shall be forwarded in a standardised format.
So clearly, we have the beginnings of a Europe wide
database of all manufacturers and their devices, which can only be of benefit
to all concerned, and in particular the manufacture's authorized representative
{or the distributor, importer or person responsible for marketing), as
appropriate.
And finally, a reiteration of the latest
interpretation of the most important definition of all;
‘authorized representative’ – means any natural or legal person established in the Community who,
explicitly designated by the manufacturer, acts and may be addressed by
authorities and bodies in the Community instead of the manufacturer, with
regard to the latter’s obligations under this directive.
Unlike some providers of AR services, we
do not make an up front or regular charge for the number of Incidents that are
reported or even Product Recalls that your Company might become involved in.
Instead, we would include our time
involved in such issues, within the Free Of Charge monthly Regulatory Support
Time, that is part of our Agreement.
We prefer to represent your interests in a
professional and honest manner at all times, charging a standard fixed All
Inclusive fee that is fixed for three years and is both fair to your company
and to ours.
If the occasion should arise where we
need to become involved in reporting an Incident or in Recalling product, then
the One Hour per Month ‘free of charge Support’ that is part of our AR agreement,
would be extended to 3 hours, in order to facilitate reporting of the Incident
or Recalling the Product, with no additional cost to your company.
Thereafter, any additional costs would be
negotiated at the time, and our special reduced hourly rates would be
applicable, as an existing client.